The effect of glycerol on signal supression during electrospray ionization analysis of proteins


Autoria(s): Mendes, M. A.; Souza, B. M.; Marques, M. R.; Palma, Mario Sergio
Contribuinte(s)

Universidade Estadual Paulista (UNESP)

Data(s)

27/05/2014

27/05/2014

04/06/2004

Resumo

The effect of salts, detergents and chaotropic agents on mass spectrometric analysis are relatively well understood, mainly due to their actions decreasing the performance of ESI interface in mass spectrometric analysis. However, there are few studies in the literature characterizing the effect of protein stabilization by glycerol, followed in some circumstances by the suppression of protein signal when ESI interface is used. The aim of the present research was to investigate in details the mass spectrometric behavior of some proteins in presence of high levels of glycerol during ESI-MS analysis. Thus, horse heart myoglobin and chicken ovalbumin were used as standard proteins. It was demonstrated that the presence of 1% (v/v) glycerol suppressed the signal of these proteins during the ESI-MS analysis, even when the sample nozzle potential was scanned from 28 to 80 V. However, when the glycerol concentration was decreased to 0.5% (v/v) and the sample cone voltage adjusted to 50 V, a perfect envelope of peaks was observed, allowing the spectrum deconvolution and the molecular mass determination with mass accuracy lower than 0.01% in each situation. A molecular explanation for this suppressive effect and for the analytical overcoming of this difficult is proposed.

Formato

339-345

Identificador

http://dx.doi.org/10.1155/2004/732938

Spectroscopy, v. 18, n. 2, p. 339-345, 2004.

0712-4813

http://hdl.handle.net/11449/67763

10.1155/2004/732938

WOS:000221869800024

2-s2.0-2442641392

2-s2.0-2442641392.pdf

Idioma(s)

eng

Relação

Spectroscopy

Direitos

openAccess

Palavras-Chave #Cone voltage #Electrospray ionization (ESI) #Signal supression #Biomedical engineering #Detergents #Ionization #Mass spectrometry #Proteins #Glycerol #glycerol #myoglobin #ovalbumin #chicken #conference paper #controlled study #electric potential #electrospray #horse #mass spectrometry #nonhuman #protein analysis #signal transduction #standardization
Tipo

info:eu-repo/semantics/conferencePaper