Endotracheal tube cuff pressure: need for precise measurement


Autoria(s): Braz, José Reinaldo Cerqueira; Navarro, Lais Helena Camacho; Takata, Ieda Harumi; Nascimento Junior, Paulo do
Contribuinte(s)

Universidade Estadual Paulista (UNESP)

Data(s)

20/05/2014

20/05/2014

04/11/1999

Resumo

Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)

CONTEXTO: Tubos traqueais com balonete de alta complacência são utilizados para prevenção de vazamento de gás e de aspiração pulmonar em pacientes submetidos à ventilação mecânica. Entretanto, o volume de insuflação habitual gera uma pressão do balonete que se transmite diretamente à parede traqueal e pode causar lesões. OBJETIVO: Testar a hipótese de que as pressões no balonete do tubo traqueal geralmente estão elevadas (acima de 40 cmH2O) na Sala de Recuperação Pós-anestésica (SRPA) ou nas Unidades de Terapia Intensiva (UTI). TIPO DE ESTUDO: Estudo de seção transversal. LOCAL: Sala de recuperação pós-anestésica e unidade de terapia intensiva. PARTICIPANTES: Medimos a pressão no balonete do tubo traqueal em 85 pacientes adultos, sendo: G1 (n=31) pacientes da UTI; G2 (n=32) pacientes da SRPA, após anestesia com óxido nitroso; G3 (n=2) pacientes da SRPA, após anestesia sem óxido nitroso. A pressão no balonete foi medida utilizando-se um manômetro (Mallinkrodt, USA). Quando necessário, retirou-se gás para ajustar a pressão no balonete até 30 cmH2O. VARIÁVEIS ESTUDADAS: Pressão do balonete do tubo traqueal. RESULTADOS: Foram observadas pressões elevadas no balonete do tubo traqueal em 90,6% dos pacientes de G2, 54,8% de G1 e 45,4% de G3 (P < 0,001). CONCLUSÕES: As pressões no balonete do tubo traqueal na UTI e na SRPA estão elevadas rotineiramente e são significativamente mais altas quando se utiliza óxido nitroso. A pressão no balonete do tubo traqueal deve ser medida rotineiramente para minimizar o trauma traqueal.

CONTEXT: High compliance endotracheal tubes cuffs are used to prevent gas leak and also pulmonary aspiration in mechanically ventilated patients. However, the use of the usual cuff inflation volumes may cause tracheal damage. OBJECTIVE: We tested the hypothesis that endotracheal tube cuff pressures are routinely high (above 40 cmH2O) in the Post Anesthesia Care Unit (PACU) or Intensive Care Units (ICU). DESIGN: Cross-sectional study. SETTING: Post anesthesia care unit and intensive care unit. PARTICIPANTS: We measured endotracheal tubes cuff pressure in 85 adult patients, as follows: G1 (n = 31) patients from the ICU; G2 (n = 32) patients from the PACU, after anesthesia with nitrous oxide; G3 (n = 22) patients from the PACU, after anesthesia without nitrous oxide. Intracuff pressure was measured using a manometer (Mallinckrodt, USA). Gas was removed as necessary to adjust cuff pressure to 30 cmH2O. MAIN MEASUREMENTS: Endotracheal tube cuff pressure. RESULTS: High cuff pressure (> 40 cmH2O) was observed in 90.6% patients of G2, 54.8% of G1 and 45.4% of G3 (P < 0.001). The volume removed from the cuff in G2 was higher than G3 (P < 0.05). CONCLUSION: Endotracheal tubes cuff pressures in ICU and PACU are routinely high and significant higher when nitrous oxide is used. Endotracheal tubes cuff pressure should be routinely measured to minimize tracheal trauma.

Formato

243-247

Identificador

http://dx.doi.org/10.1590/S1516-31801999000600004

São Paulo Medical Journal. Associação Paulista de Medicina - APM, v. 117, n. 6, p. 243-247, 1999.

1516-3180

http://hdl.handle.net/11449/30608

10.1590/S1516-31801999000600004

S1516-31801999000600004

2-s2.0-0033523796

S1516-31801999000600004.pdf

Idioma(s)

eng

Publicador

Associação Paulista de Medicina (APM)

Relação

São Paulo Medical Journal

Direitos

openAccess

Palavras-Chave #Tubo traqueal #Pressão do balonete #Óxido nitroso #Tracheal tube #Intracuff pressure #Nitrous oxide
Tipo

info:eu-repo/semantics/article