Determination of Zinc in Lubricating Oil by Flame MS Employing Ultrasonic Extraction


Autoria(s): Sabio, Rafael Miguel; Oliveira, Silvana Ruella; Tognolli, Joao Olimpio; Gomes Neto, José Anchieta
Contribuinte(s)

Universidade Estadual Paulista (UNESP)

Data(s)

20/05/2014

20/05/2014

01/11/2011

Resumo

Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)

Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)

The determination of zinc in lubricating oils by line-source flame atomic absorption spectrometry (LS-FAAS) is proposed employing an alternative sample preparation procedure based on ultrasonic wave-assisted acid extraction. The chemometric optimization of the main parameters influencing the extraction provided the following conditions: 1.3 mL of HCl, 5.0 mL of H(2)O(2), 120 seconds sonication time, 60 degrees C heating temperature, and 45% sonication amplitude of the ultrasound probe. Analytical curves in the 0.5 - 4.0 mg L(-1) Zn range were consistently obtained (R = 0.9990). Commercial unused lubricating oil samples were analyzed by the proposed method and by high-resolution continuum source flame AAS (HR-CS FAAS) as the comparative technique. A paired t-test showed that the results between the two techniques were in agreement at the 95% confidence level. The accuracy of the proposed method was evaluated using an alternative sample preparation method based on recovery tests. Recoveries in the 92 - 119% interval were found. The relative standard deviation was 4% and the limit of detection was 17.6 mu g L(-1).

Formato

240-245

Identificador

http://www.lw20.com/1201301026164566.html

Atomic Spectroscopy. Shelton: Perkin-elmer Corp, v. 32, n. 6, p. 240-245, 2011.

0195-5373

http://hdl.handle.net/11449/25761

WOS:000298114900005

Idioma(s)

eng

Publicador

Perkin-elmer Corp

Relação

Atomic Spectroscopy

Direitos

closedAccess

Tipo

info:eu-repo/semantics/article