Symmetrically tapered <30-micron-thick quasi-planar Ge waveguides as chemical sensors for microanalysis


Autoria(s): Vongsvivut, Jitraporn; Shilov, Sergey V.; Ekgasit, Sanong; Braiman, Mark S.
Data(s)

01/12/2002

Resumo

Symmetrically tapered planar IR waveguides have been fabricated by starting with a ZnS coated concave piece of single-crystal Ge, embedding it in an epoxide resin as a supporting substrate, and then grinding and polishing a planar surface until the thickness at the taper minimum is <30 μm. Such tapering is expected to enhance a waveguide's sensitivity as an evanescent wave sensor by maximizing the amount of evanescent wave energy present at the thinnest part of the waveguide. As predicted by theory, the surface sensitivity, i.e., the absorbance signal per molecule in contact with the sensing region, increases with decreasing thickness of the tapered region even while the total energy throughput decreases. The signal-to-noise ratio obtained depends very strongly on the quality of the polished surfaces of the waveguides. The surface sensitivity is superior to that obtained with a commercial Ge attenuated total reflection (ATR) accessory for several types of sample, including thin films (<10 ng) and small volumes (<1 μL) of volatile solvents. By using the waveguides, light-induced structural changes in the protein bacteriorhodopsin were observable using samples as small as ∼50 pmol (∼1 μg). In addition, the waveguide sensors can reveal the surface compositions on a single human hair, pointing to their promise as a tool for forensic fiber analysis.

Identificador

http://hdl.handle.net/10536/DRO/DU:30038962

Idioma(s)

eng

Publicador

Society for Applied Spectroscopy

Relação

http://dx.doi.org/10.1366/000370202321116011

Palavras-Chave #bacteriorhodopsin (bR) #D96N mutant #evanescent wave sensor #hair analysis #nujol #sensitivity #tapered Ge waveguides #throughput
Tipo

Journal Article