Square wave adsorptive cathodic stripping voltarnmetry automated by sequential injection analysis - Potentialities and limitations exemplified by the determination of methyl parathion in water samples


Autoria(s): SANTOS, Luciana B. O. dos; MASINI, Jorge C.
Contribuinte(s)

UNIVERSIDADE DE SÃO PAULO

Data(s)

20/10/2012

20/10/2012

2008

Resumo

This paper describes the development and evaluation of a sequential injection method to automate the determination of methyl parathion by square wave adsorptive cathodic stripping voltammetry exploiting the concept of monosegmented flow analysis to perform in-line sample conditioning and standard addition. Accumulation and stripping steps are made in the sample medium conditioned with 40 mmol L-1 Britton-Robinson buffer (pH 10) in 0.25 mol L-1 NaNO3. The homogenized mixture is injected at a flow rate of 10 mu Ls(-1) toward the flow cell, which is adapted to the capillary of a hanging drop mercury electrode. After a suitable deposition time, the flow is stopped and the potential is scanned from -0.3 to -1.0 V versus Ag/AgCl at frequency of 250 Hz and pulse height of 25 mV The linear dynamic range is observed for methyl parathion concentrations between 0.010 and 0.50 mgL(-1), with detection and quantification limits of 2 and 7 mu gL(-1), respectively. The sampling throughput is 25 h(-1) if the in line standard addition and sample conditioning protocols are followed, but this frequency can be increased up to 61 h(-1) if the sample is conditioned off-line and quantified using an external calibration curve. The method was applied for determination of methyl parathion in spiked water samples and the accuracy was evaluated either by comparison to high performance liquid chromatography with UV detection, or by the recovery percentages. Although no evidences of statistically significant differences were observed between the expected and obtained concentrations, because of the susceptibility of the method to interference by other pesticides (e.g., parathion, dichlorvos) and natural organic matter (e.g., fulvic and humic acids), isolation of the analyte may be required when more complex sample matrices are encountered. (C) 2007 Elsevier B.V. All rights reserved.

Identificador

ANALYTICA CHIMICA ACTA, v.606, n.2, p.209-216, 2008

0003-2670

http://producao.usp.br/handle/BDPI/31351

10.1016/j.aca.2007.11.009

http://dx.doi.org/10.1016/j.aca.2007.11.009

Idioma(s)

eng

Publicador

ELSEVIER SCIENCE BV

Relação

Analytica Chimica Acta

Direitos

restrictedAccess

Copyright ELSEVIER SCIENCE BV

Palavras-Chave #methyl parathion #adsorptive voltammetry #sequential injection #waters #MERCURY DROP ELECTRODE #ORGANOPHOSPHORUS PESTICIDES #NATURAL-WATERS #LIQUID-CHROMATOGRAPHY #GAS-CHROMATOGRAPHY #MASS-SPECTROMETRY #NERVE AGENTS #VOLTAMMETRY #SYSTEM #INSECTICIDES #Chemistry, Analytical
Tipo

article

original article