Inter-observer agreement in evaluating the anterior chamber angle by ultrasound biomicroscopy


Autoria(s): Azuara-Blanco, A.; Spaeth, G.L.; Araujo, S.V.; Augsburger, J.J.
Data(s)

15/02/1996

Resumo

Purpose. To determine the agreement between observers in estimating the configuration of the human anterior chamber angle (ACA) using ultrasound biomicroscopy (UBM). Methods. Two masked clinicians used UBM images to estimate, in 41 eyes, the configuration of the ACA, especially (1) the position of contact between the peripheral iris and the inside of the eye wall, (2) the angularity of the approach to the ACA, and (3) the curvature of the peripheral iris. Agreement between observers was evaluated by the kappa statistic. Results. Inter-observer agreement in assessing the iris insertion (kappa = 0.79), angular width (Kappa = 0.95), and the peripheral iris curvature (kappa = 0.84) was high. Conclusions. The agreement between observers in evaluating the anterior chamber angle configuration by UBM was excellent.

Identificador

http://pure.qub.ac.uk/portal/en/publications/interobserver-agreement-in-evaluating-the-anterior-chamber-angle-by-ultrasound-biomicroscopy(8330def4-dd21-4373-a677-5343c47ff2e1).html

http://www.scopus.com/inward/record.url?partnerID=yv4JPVwI&eid=2-s2.0-33750197961&md5=10076aa5fd7f195436b668e71dd363a5

Idioma(s)

eng

Direitos

info:eu-repo/semantics/restrictedAccess

Fonte

Azuara-Blanco , A , Spaeth , G L , Araujo , S V & Augsburger , J J 1996 , ' Inter-observer agreement in evaluating the anterior chamber angle by ultrasound biomicroscopy ' Investigative ophthalmology & visual science , vol 37 , no. 3 .

Tipo

article