Error recovery in cyberphysical digital microfluidic biochips


Autoria(s): Luo, Y; Chakrabarty, K; Ho, TY
Data(s)

07/01/2013

Formato

59 - 72

Identificador

IEEE Transactions on Computer-Aided Design of Integrated Circuits and Systems, 2013, 32 (1), pp. 59 - 72

0278-0070

http://hdl.handle.net/10161/5741

http://hdl.handle.net/10161/5741

Idioma(s)

en_US

Relação

IEEE Transactions on Computer-Aided Design of Integrated Circuits and Systems

ECE;2012-3

10.1109/TCAD.2012.2211104

Palavras-Chave #Biochips #cyberphysical systems #microfluidics
Tipo

Journal Article

Resumo

Droplet-based digital microfluidics technology has now come of age, and software-controlled biochips for healthcare applications are starting to emerge. However, today's digital microfluidic biochips suffer from the drawback that there is no feedback to the control software from the underlying hardware platform. Due to the lack of precision inherent in biochemical experiments, errors are likely during droplet manipulation; error recovery based on the repetition of experiments leads to wastage of expensive reagents and hard-to-prepare samples. By exploiting recent advances in the integration of optical detectors (sensors) into a digital microfluidics biochip, we present a physical-aware system reconfiguration technique that uses sensor data at intermediate checkpoints to dynamically reconfigure the biochip. A cyberphysical resynthesis technique is used to recompute electrode-actuation sequences, thereby deriving new schedules, module placement, and droplet routing pathways, with minimum impact on the time-to-response. © 2012 IEEE.