Amperometric glucose biosensor based on lipid film


Autoria(s): Wu ZY; Wang BQ; Dong SJ; Wang EK
Data(s)

2000

Resumo

A novel glucose biosensor based on cast lipid film was developed. This model of biological membrane was used to supply a biological environment on the surface of the electrode, moreover it could greatly reduce the interference and effectively exclude hydrophilic electroactive material from reaching the detecting surface. TTF was selected as a mediator because of its high electron-transfer efficiency, and it was incorporated in the lipid film firmly. Glucose oxidase was immobilized in hydrogel covered on the lipid film. The effects of pH, operating potential were explored for the optimum analytical performance by using amperometric method. The response time of the biosensor was less than 20 s, and the linear range is up to 10 mmol l(-1) (corr. coeff. 0.9932) with the detection limit of 2 x 10(-5) mol l(-1). The biosensor also exihibited good stability and reproducibility. (C) 2000 Elsevier Science S.A. All rights reserved.

Identificador

http://202.98.16.49/handle/322003/19677

http://www.irgrid.ac.cn/handle/1471x/154573

Idioma(s)

英语

Fonte

Wu ZY;Wang BQ;Dong SJ;Wang EK.Amperometric glucose biosensor based on lipid film,BIOSENSORS & BIOELECTRONICS,2000,15(3-4 ):143-147

Palavras-Chave #ELECTRON-TRANSFER #ENZYME ELECTRODE #PHASE-TRANSITION #CONSTRUCTION #INJECTION #MEMBRANES
Tipo

期刊论文