Resumo

The determination of glucose is possible with the enzymatic reaction of glucose oxidase and potentiometric detection. The signal is proportional to the concentration up to 50 mg/dl. This value is fixed by the concentration of oxygen in the sample. By adding catalase, concentrations up to 2000 mg/dl are detectable. The steepness of the calibration curve is not affected by oxygen concentrations greater than 4 mg/l. In contrast to amperometric sensors, an influence of deposits on the electrodes surface on the signal cannot be found with potentiometric sensors

Identificador

http://ir.yic.ac.cn/handle/133337/3010

http://www.irgrid.ac.cn/handle/1471x/139946

Fonte

Honold F; Cammann K..Potentiometric glucose sensors,Hormone and Metabolic Research Supplement Series,1988,20(47):9

Palavras-Chave #Biosensing Techniques #Catalase #Electrodes #Glucose/analysis* #Glucose Oxidase #Hydrogen-Ion Concentration #Oxygen/pharmacology #Potentiometry