The application of spline wavelet transform to the determination of electroosmotic mobility in capillary electrophoresis


Autoria(s): Xiong, JH; Xu, GW; Zhang, PD; Zheng, YF; Zhang, YK
Data(s)

19/07/2002

Resumo

The rule of current change was studied during capillary electrophoresis (CE) separation process while the conductivity of the sample solution was different from that of the buffer. Using a quadratic spline wavelet of compact support, the wavelet transforms (WTs) of capillary electrophoretic currents were performed. The time corresponding to the maximum of WT coefficients was chosen as the time of current inflection to calculate electroosmotic mobility. The proposed method was suitable for different CE modes, including capillary zone electrophoresis, nonaqueous CE and micellar electrokinctic chromatography. Compared with the neutral marker method, the relative errors of the developed method for the determination of electroosmotic mobility were all below 2.5%. (C) 2002 Elsevier Science B.V. All rights reserved.

Identificador

http://159.226.238.44/handle/321008/83183

http://www.irgrid.ac.cn/handle/1471x/138850

Idioma(s)

英语

Fonte

熊建辉;许国旺;张普敦;郑育芳;张玉奎.The application of spline wavelet transform to the determination of electroosmotic mobility in capillary electrophoresis,Talanta,2002,57(6):1093-1100

Palavras-Chave #capillary electrophoresis #electroosmotic mobility #spline wavelet transform
Tipo

期刊论文