Design, simulation and testing of large area silicon drift detectors and detector array for X-ray spectroscopy


Autoria(s): Zhang WC; Li Z; Siddons DP; Huang T; Zhao LJ; Kakuno EM; Pietraski P; Li CJ
Data(s)

2000

Resumo

Large area (25 mm(2)) silicon drift detectors and detector arrays (5x5) have been designed, simulated, and fabricated for X-ray spectroscopy. On the anode side, the hexagonal drift detector was designed with self-biasing spiral cathode rings (p(+)) of fixed resistance between rings and with a grounded guard anode to separate surface current from the anode current. Two designs have been used for the P-side: symmetric self-biasing spiral cathode rings (p(+)) and a uniform backside p(+) implant. Only 3 to 5 electrodes are needed to bias the detector plus an anode for signal collection. With graded electrical potential, a sub-nanoamper anode current, and a very small anode capacitance, an initial FWHM of 1.3 keV, without optimization of all parameters, has been obtained for 5.9 keV Fe-55 X-ray at RT using a uniform backside detector.

Identificador

http://ir.semi.ac.cn/handle/172111/12428

http://www.irgrid.ac.cn/handle/1471x/65184

Idioma(s)

英语

Fonte

Zhang WC; Li Z; Siddons DP; Huang T; Zhao LJ; Kakuno EM; Pietraski P; Li CJ .Design, simulation and testing of large area silicon drift detectors and detector array for X-ray spectroscopy ,IEEE TRANSACTIONS ON NUCLEAR SCIENCE,2000,47(4 Part.1.):1381-1385

Palavras-Chave #半导体器件
Tipo

期刊论文