Toward an understanding of the turbidity measurement of fleterocoagulation rate constants of dispersions containing particles of different sizes


Autoria(s): 刘捷; 徐升华; 孙祉伟
Data(s)

2007

Resumo

Our previous studies have shown that the determination of coagulation rate constants by turbidity measurement becomes impossible for a certain operating wavelength (that is, its blind point) because at this wavelength the change in the turbidity of a dispersion completely loses its response to the coagulation process. Therefore, performing the turbidity measurement in the wavelength range near the blind point should be avoided. In this article, we demonstrate that the turbidity measurement of the rate constant for coagulation of a binary dispersion containing particles of two different sizes (heterocoagulation) presents special difficulties because the blind point shifts with not only particle size but also with the component fraction. Some important aspects of the turbidity measurement for the heterocoagulation rate constant are discussed and experimentally tested. It is emphasized that the T-matrix method can be used to correctly evaluate extinction cross sections of doublets formed during the heterocoagulation process, which is the key data determining the rate constant from the turbidity measurement, and choosing the appropriate operating wavelength and component fraction are important to achieving a more accurate rate constant. Finally, a simple scheme in experimentally determining the sensitivity of the turbidity changes with coagulation over a wavelength range is proposed.

Identificador

http://dspace.imech.ac.cn/handle/311007/33875

http://www.irgrid.ac.cn/handle/1471x/2786

Idioma(s)

英语

Fonte

Langmuir.2007,23(23):11451-11457

Palavras-Chave #Heteroaggregation Rate Constants #Dynamic Light-Scattering #Polymer Colloids #Heterocoagulation #Coagulation #Homocoagulation #Aggregation #Stability #Kinetics #Latex
Tipo

期刊论文