Passive control of a biventricular assist device with compliant inflow cannulae


Autoria(s): Gregory, Shaun D.; Pearcy, Mark J.; Timms, Daniel L.
Data(s)

01/08/2012

Resumo

Rotary ventricular assist device (VAD) support of the cardiovascular system is susceptible to suction events due to the limited preload sensitivity of these devices. This may be of particular concern with rotary biventricular support (BiVAD) where the native, flow-balancing Starling response is diminished in both ventricles. The reliability of sensor and sensor-less based control systems which aim to control VAD flow based on preload have limitations and thus an alternative solution is desired. This study introduces a compliant inflow cannula (CIC) which could improve the preload sensitivity of a rotary VAD by passively altering VAD flow depending on preload. To evaluate the design, both the CIC and a standard rigid inflow cannula were inserted into a mock circulation loop to enable biventricular heart failure support using configurations of atrial and ventricular inflow, and arterial outflow cannulation. A range of left (LVAD) and right VAD (RVAD) rotational speeds were tested as well as step changes in systemic/pulmonary vascular resistance to alter relative preloads, with resulting flow rates recorded. Simulated suction events were observed, particularly at higher VAD speeds, during support with the rigid inflow cannula, while the CIC prevented suction events under all circumstances. The compliant section passively restricted its internal diameter as preload was reduced, which increased the VAD circuit resistance and thus reduced VAD flow. Therefore, a compliant inflow cannula could potentially be used as a passive control system to prevent suction events in rotary left, right and biventricular support.

Formato

application/pdf

Identificador

http://eprints.qut.edu.au/63961/

Publicador

Wiley-Blackwell Publishing, Inc.

Relação

http://eprints.qut.edu.au/63961/1/Passive_Control_Artificial_Organs.pdf

DOI:10.1111/j.1525-1594.2012.01504.x

Gregory, Shaun D., Pearcy, Mark J., & Timms, Daniel L. (2012) Passive control of a biventricular assist device with compliant inflow cannulae. Artificial Organs, 36(8), pp. 683-690.

Direitos

© 2012, Copyright the Authors. Artificial Organs © 2012, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc

The definitive version is available at www3.interscience.wiley.com

Fonte

School of Chemistry, Physics & Mechanical Engineering; Institute of Health and Biomedical Innovation; Science & Engineering Faculty

Palavras-Chave #090300 BIOMEDICAL ENGINEERING #090302 Biomechanical Engineering #090304 Medical Devices #110201 Cardiology (incl. Cardiovascular Diseases) #Artificial heart #Compliant cannula #suction event #physiological control #biventricular assist device #heart failure
Tipo

Journal Article