Reconstructing 3D x-ray CT images of polymer gel dosimeters using the zero-scan method


Autoria(s): Kakakhel, Muhammad; Kairn, Tanya; Kenny, John; Trapp, Jamie
Data(s)

2013

Resumo

In this study x-ray CT has been used to produce a 3D image of an irradiated PAGAT gel sample, with noise-reduction achieved using the ‘zero-scan’ method. The gel was repeatedly CT scanned and a linear fit to the varying Hounsfield unit of each pixel in the 3D volume was evaluated across the repeated scans, allowing a zero-scan extrapolation of the image to be obtained. To minimise heating of the CT scanner’s x-ray tube, this study used a large slice thickness (1 cm), to provide image slices across the irradiated region of the gel, and a relatively small number of CT scans (63), to extrapolate the zero-scan image. The resulting set of transverse images shows reduced noise compared to images from the initial CT scan of the gel, without being degraded by the additional radiation dose delivered to the gel during the repeated scanning. The full, 3D image of the gel has a low spatial resolution in the longitudinal direction, due to the selected scan parameters. Nonetheless, important features of the dose distribution are apparent in the 3D x-ray CT scan of the gel. The results of this study demonstrate that the zero-scan extrapolation method can be applied to the reconstruction of multiple x-ray CT slices, to provide useful 2D and 3D images of irradiated dosimetry gels.

Formato

application/pdf

Identificador

http://eprints.qut.edu.au/61031/

Publicador

Institute of Physics Publishing Ltd.

Relação

http://eprints.qut.edu.au/61031/1/Dosgel_CT_Paper_-_Kakakhel_et_al.pdf

DOI:10.1088/1742-6596/444/1/012091

Kakakhel, Muhammad, Kairn, Tanya, Kenny, John, & Trapp, Jamie (2013) Reconstructing 3D x-ray CT images of polymer gel dosimeters using the zero-scan method. Journal of Physics: Conference Series, 444, 012091.

Direitos

Copyright 2013 (please consult the author).

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. Published under licence by IOP Publishing Ltd

Fonte

School of Chemistry, Physics & Mechanical Engineering; Science & Engineering Faculty

Palavras-Chave #029903 Medical Physics #gel dosimetry #ct #zero scan