Magnitude and variability of loading on the osseointegrated implant of transfemoral amputees during walking


Autoria(s): Lee, Winson C.C.; Frossard, Laurent A.; Hagberg, Kerstin; Haggstrom, Eva; Lee Gow , David; Gray , Steven
Data(s)

2008

Resumo

This study directly measured the load acting on the abutment of the osseointegrated implant system of transfemoral amputees during level walking, and studied the variability of the load within and among amputees. Twelve active transfemoral amputees (age: 54±12 years, mass:84.3±16.3 kg, height: 17.8±0.10 m) fitted with an osseointegrated implant for over 1 year participated in the study. The load applied on the abutment was measured during unimpeded, level walking in a straight line using a commercial six-channel transducer mounted between the abutment and the prosthetic knee. The pattern and the magnitude of the three-dimensional forces and moments were revealed. Results showed a low step-to-step variability of each subject, but a high subject-to-subject variability in local extrema of body-weight normalized forces and moments and impulse data. The high subject-to-subject variability suggests that the mechanical design of the implant system should be customized for each individual, or that a fit-all design should take into consideration the highest values of load within a broad range of amputees. It also suggests specific loading regime in rehabilitation training are necessary for a given subject. Thus the loading magnitude and variability demonstrated should be useful in designing an osseointegrated implant system better able to resist mechanical failure and in refining the rehabilitation protocol.

Formato

application/pdf

Identificador

http://eprints.qut.edu.au/29105/

Publicador

Elsevier Ltd

Relação

http://eprints.qut.edu.au/29105/1/c29105.pdf

DOI:10.1016/j.medengphy.2007.09.003

Lee, Winson C.C., Frossard, Laurent A., Hagberg, Kerstin, Haggstrom, Eva, Lee Gow , David, & Gray , Steven (2008) Magnitude and variability of loading on the osseointegrated implant of transfemoral amputees during walking. Medical Engineering & Physics, 30(7), pp. 825-833.

Direitos

Copyright 2007 IPEM Published by Elsevier Ltd.

Fonte

Institute of Health and Biomedical Innovation

Palavras-Chave #110699 Human Movement and Sports Science not elsewhere classified #Transfemoral amputation #Prosthetics #Osseointegration #Transducer #Gait
Tipo

Journal Article